Download Hemolyzer - Index of
Transcript
69 Service Manual 7.2.2. Database relating warnings Code Reason(s) Message 5101 %u data selected. 5106 No such type of data 5108 Connect an USB storage This message device with data! appears at View external function. Connect storage media into the USB port and press . 5109 % data record(s) will be The analyzer will saved on an USB save %u number of storage device. data records. Insert a storage device into the USB port. Insert USB storage media and press . record(s) is There are %u data records present in the database which match the selection. Remedy Press There are no %u Press data records present selection. in the database which match the selection. . and check the 7.2.3. Warnings relating to QC measure Code 5200 Message Reason(s) Your previous QC data When you press will be lost. for accepting Are you sure? new QC target values. Remedy If you press the new values will be accepted and all of the measurements on the selected QC level will be deleted. If you press old values remain 7.2.4. Calibration relating warnings Code 5300 Message Reason(s) Some or all of the During automatic calibration factors are out calibration at least of range! one of the factors are out of the range 0.8 or 1.2. Remedy Check the target values and the control blood and repeat the calibration. If the problem still persists contact the Service Personnel.