Download Agilent Technologies 1650B/1651B Logic Analyzer Service Manual
Transcript
Removal and Replacement of Power Supply 1. Disconnect pod cables from rear of instrument. 2. Disconnect power cable. 3. Remove ten screws securing top cover. Remove top cover. 4. Disconnect disk drive cable from disk drive. 5. Remove two screws securing disk drive to power supply. Remove disk drive by sliding assembly out front panel of instrument. 6. Remove two retaining pins securing power supply to chassis. Slide power supply out far enough to gain unobstructed access to power supply cables. 7. Disconnect power supply line input and DC output cables. Remove power supply. 8. Slide new power supply in far enough to connect line input cable and DC output cable. Connect cables and slide power supply in completely. Secure power supply using two retaining pins. 9. Slide disk drive through front panel of instrument. Secure disk drive to power supply using two screws. 10. Connect disk drive cable to disk drive. 11. Install top cover. Secure top cover to chassis using ten screws. Note Removal and Replacement of Fan To ensure top cover is properly installed, secure side screws of top cover before securing top screws. 1. Disconnect pod cables from rear of instrument. 2. Disconnect power cable. 3. Remove ten screws securing top cover. Remove top cover. 4. Disconnect fan cable and BNC trigger cables from system board. 5. Disconnect AC power supply cable from power supply. 6. Remove two hex standoffs securing HP-IB cable to rear panel. 7. Remove eight screws securing rear panel. Remove rear panel. Note An ESD spring ground is inserted on the RS-232 connector. This spring ground is not secured and can easily fall off. 8. Detach fan from rear panel by removing four screws securing fan to rear panel. 9. Disconnect fan cable from old fan and connect to new fan. Service 6-20 HP 1650B/1651B Service Manual