Download Manual Part Number - Artisan Technology Group
Transcript
Notes Example When repeating the procedure, all references to channel 1 should be changed to the channel being tested. Turn off all unused channels using the BLANk command. Bandwidth errors can be caused by the need for adjustment. Perform high frequency pulse response a&ustment, (see Adjustment procedures, Chapter 5) before troubleshooting instrument. If adjustment fails to correct problem, cause may be the attenuator or A2 PCA See troubleshooting in Chapter 8. This program is provided to show how to perform a bandwidth test, and print out the results of each measurement. The program pauses to allow the user to a&ust the signal generator and power meter. The example is written using: an HP-IB select code of 7, primary address of 09, and secondary address of 05 for the HP E1426A COMPatible programming language an HP Series 200/300 Computer with HP BASIC ecut OUTPUT 70905; "*CLSm OUTPUT 70905; "'RST" OUTPUT 70905;"TIM:RANG 2E-6" OUTPUT 70905; "VIEW CHAN1" OUTPUT 70905;"BLAN CHAN2, CHAN3, CHAN4" OUTPUT 70905;"CHANl:COUP DCF" OUTPUT 70905; *'TRIG:SOUR CHANl" OUTPUT 70905;"ACQ:TYP AVER" OUTPUT 70905; "ACQ:COUN 32" OUTPUT 70905;"CHANl:RANG 3.2E-1" OUTPUT 70905;"MEAS:SOUR CHAN1" OUTPUT 70905; "DIG CHAN1" OUTPUT 70905; "MEAS:VPP?'* ENTER 70905;A PRINT A ! To adjust signal generator and power meter. PAUSE OUTPUT 70905;"TIM:RANG 5E-9" OUTPUT 70905;"MEAS:SOUR CHAN1" OUTPUT 70905;"DIG CHAN1" OUTPUT 70905;"MEAS:VPP?" ENTER 70905;B PRINT B PAUSE ! To calculate response and change test set up. ! Repeat ! Repeat for channels 2-4 . for <80 m V range. END February 9, 1991 Verification Tests 4-17