Reference
Keys to Using
Read the following topics; they provide details that can help set up to take
effective differential and common mode TDR measurements.
The 80E04 TDR/sampling module is able to perform differential and commonmode TDR measurements. As described earlier, the sampling module has two
input channels and two independent step generators.
The step-generator output for each channel is selectable for positive or negative
polarity and amplitude. This section will show you how to use the two channels
and step generators of the 80E04 to perform differential and common-mode
TDR measurements.
To Take a Common-Mode
or Differential TDR
Measurement
Overview
This example demonstrates the common-mode and differential TDR features of
the 80E04 sampling module.
To take a common mode or differential TDR
measurement
Prerequisites 1.
2.
Connect your wrist strap to the antistatic connector on
the front of your instrument.
Control elements & resources
Connect
wrist strap
An 80E04 sampling module must be installed in a
TDS oscilloscope or CSA analyzer. The acquisition
system should be set to Run.
See the main instrument user documentaion for
scaling and acquisition setup
Input 3.
Connect transmission lines to the sampling module
using proper probing/connecting techniques for your
application (for example: two SMA cables, preferably of
matched length). Connect the device under test to the
transmission lines (Connect the conductors of a
differential line to the center conductors. Connect the
shields together.)
80E00 Electrical Sampling Modules User Manual
33