Download CSA7404B, TDS7704B, TDS7404
Transcript
Brief Procedures H As you rotate the lower multipurpose knob, the square-wave probe-compensation signal can become triggered and untriggered. (Leave the signal triggered.) c. Verify the delayed trigger counter: H Double click the Trig Delay control to pop up a numeric keypad for that control. H Click the keypad to enter a trigger delay time of 1 second and then click Enter. H Verify that the trigger READY indicator on the front panel flashes about once every second as the waveform is updated on-screen. 8. Remove the test hookup: Disconnect the BNC cable from the channel input and the probe compensation output. Verify the File System Equipment required One BNC cable, such as Tektronix part number 012-0076-00 One TCA-BNC adapter One 1.44 Mbyte, 3.5 inch DOS-compatible formatted disk. Prerequisites None 1. Initialize the instrument:Push the front-panel DEFAULT SETUP button. 2. Hook up the signal source: Connect the BNC cable from the probe compensation output to the CH 1 input through a TCA-BNC adapter as shown in Figure 4--6. Instrument under test BNC cable from PROBE COMPENSATION output to CH 1 input Figure 4- 6: Setup for the file system test 3. Insert the test disk: Insert the floppy disk in the floppy disk drive. 4. Set up the instrument: Push the front panel AUTOSET button. 4- 14 CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual