Download - Frank`s Hospital Workshop

Transcript
Capnometry (etCO2)
altitude error
0.0%
500
1000
1500
2000
2500
3000
Displaying the Capnogram
3500
4000
4500
5000
-5.0%
The capnogram can be displayed in channel 3 with
a sweep speed of 6.25 mm/s (but it is printed with
a speed of 25 mm/s).
-10.0%
error
-15.0%
-20.0%
•
•
•
-25.0%
-30.0%
-35.0%
-40.0%
altitude (m)
Figure 10-6. Measurement accuracy as a function
of the altitude of the measuring site
•
↓
ECG
QRSPulse
Tone
OFF
SpO2
etCO2
Next
Menu
Figure 10-7. Main menu
↓
Assign
Channel
Waveform
Filter
ON
Memory
Display
ECG
Vit. Sign
Previous
Menu
Figure 10-9. Menu to assign waveforms to
channels
89
160 / 40
ECG
(Figure 10-7).
Press F2
Assign Channel Waveform
(Figure 10-8).
Press F3 (Figure 10-9) repeatedly until "etCO2"
is displayed (Figure 10-10).
When the etCO2 measurement is disabled, a
sawtooth signal is displayed which can be
replaced with an ECG lead.
During the automatic sensor calibration
process, a calibration pulse is superimposed on
the capnogram (a, Figure 10-10).
Cleaning the Sensor
↓
ECG
Vit. Sign
Next Menu
Previous
Menu
Figure 10-8. Main menu, page 2
2 Channel 3
Channel 1 Channel
ECG
•
Press F5
Use a soft cloth moistened with an alcoholic, nonabrasive cleaning agent to wipe the sensor surface
including the IR light source, the photodetector
and the connecting lead carefully clean. Use an
alcoholic, non-abrasive cleaning agent.
I
Caution
bpm
Damage to Sensor — Do not use caustic or abraII
37
etCO2
a
--20
mmHg
etCO2
sive cleaners or hard cleaning utensils.
Do not touch the surface of the IR light source or
of the photodetector with fingers or any hard
object. This could affect the performance of these
15.07.1999
fine optical devices.
09:05:00
semiautom.
ECG
etCO2
SpO2
QRSPulse
Tone
OFF
Next
Menu
Figure 10-10.Capnogram in channel 3
a automatic sensor calibration
Do not autoclave the sensor or sterilize it with
ethylene oxide.
Do not immerse the sensor in liquid.
Warning
Infection Hazard — The airway adapter is for
single use only and must not be reused.
227 490 02-C
Marquette Responder® 3000
57