ECG: ECG Menu
Display Lead
This option allows you to change the lead displayed as the primary
lead—the one in the top trace position. This is useful when you are
selecting a lead for pacemaker detection or analog output.
Select the DISPLAY LEAD option from the ECG menu. A popup menu
opens, showing all leads available. The lead currently being displayed is
highlighted.
127(
There is only one V lead choice, even if you are using a 10-leadwire
cable. The V lead available is determined by the V lead identified in
the ST menu.
If you are using a standard (selectable lead), three-leadwire cable,
the available choices in the popup menu are leads I, II, and III.
Highlight the desired lead. The change occurs immediately on the screen.
When you are satisfied with the lead selection, press the Trim Knob
control or touch the DISPLAY LEAD option to close the popup menu.
Synchronized Cardioversion
The lead displayed in the top trace position is the signal output to the
DEFIB SYNC connector on the front of the module. The software
provides a defib sync pulse for digital synchronization and places a
return marker on the ECG waveform.
If the defibrillator is to be synchronized with the analog ECG signal,
review the patient’s ECG leads and place the one with the greatest
amplitude in the top position on the monitor.
11-20
Solar 8000M Patient Monitor
2000701-067
Revision B