GE MEDICAL SYSTEMS
DIRECTION 2286865, REVISION 5
10-5-2-2
LOGIQ™ 7 SERVICE MANUAL
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted. Refer
to the User Manual for a list of approved peripherals/options.
Table 10-6
10-5-3
10-5-3-1
Step
Item
1
VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3
Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4
DICOM
5
InSite/iLinq
6
Camera
7
Footswitch
8
ECG
9
3D Probe
Verify that DICOM is functioning properly. Send an image to a DICOM device.
Verify that InSite is functioning properly. Ensure two-way remote communications.
(Warranty & Contract Customers only)
Verify hardcopy output of the film camera. Clean as necessary.
Verify that the footswitch is functioning as programed. Clean as necessary.
Verify basic operation with customer
Input Power
Mains Cable Inspection
Step
1
10-5-4-1
Description
2
Table 10-7
10-5-4
GE Approved Peripheral/Hardware Option Functional Checks
Mains Cable Inspection
Item
Description
Unplug Cord Disconnect the mains cable from the wall and system.
2
Inspect
3
Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no
strands may cause a short circuit.
4
Verify
Inlet connector retainer is functional.
Inspect it and its connectors for damage of any kinds.
Cleaning
General Cleaning
Table 10-8
General Cleaning
Step
Item
1
Console
Description
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
2
Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
3
Monitor and
Touch Panel
Chapter 10 -
10-7