2. SimCube Specifications
Physical Dimensions
Size
Weight
Power
NIBP Connection
ECG/Resp Connection
IBP Connection
Manometer
Range
Precision
Accuracy
User Interface
Single Button Operation
Operating Modes
NIBP Adult Simulation
Simulated Pressure
Simulated Heart Rate
Simulated Pulse Volume
NIBP Neonatal Simulation
Simulated Pressure
Simulated Heart Rate
Simulated Pulse Volume
NIBP Hypertensive Simulation
Simulated Pressure
Simulated Heart Rate
Simulated Pulse Volume
NIBP Hypotensive Simulation (SC-5)
Simulated Pressure
Simulated Heart Rate
Simulated Pulse Volume
ECG Simulation (SC-2, 4, 5)
Isolated
Synchronized with NIBP
R Wave Size
R Wave Width
Wave Shape
3”x 3” x 3.5” (7.6cm X 7.6cm X 8.9cm)
2.5 Lbs
External A/C Adaptor
(Output: 6VDC / 2amps,
2.1mm, center positive connector) or
4 AA Batteries (with Battery Boost Option)
Quick Disconnect, Female
10 ECG snaps
Mini-DIN
(SC-1, 2, 3, 4) 0 – 480 mmHg
(SC-5) – 400 to + 400 mmHg
(SC-1, 2, 3, 4) 0.5 mmHg
(SC-5) 0.1 mmHg
+/-1% of reading
Adult NIBP
Neo NIBP
Hypertensive NIBP
Hypotensive NIBP (SC-5 only)
Manometer
Peak Detect (SC-3, 4, 5 only)
HR Seq. Alarm Test (SC-4, 5 only)
ECG Pace ON
Arrhythmia Sequence (SC-5 only)
Heart Rate Sequence (SC-4, 5 only)
Invasive BP Zero (SC-5 only)
Invasive BP 100, 200 (SC-5 only)
Invasive BP Sequence (SC-5 only)
120/80 (100) mmHg
70 bpm
1 ml
70/40 (55) mmHg
95 bpm
0.5 ml
190/120 (150) mmHg
70 bpm
1 ml
80/40 (60) mmHg
70 bpm
1 ml
Yes
Yes
1mV (lead I) +/- 5%
35 ms
QRS wave