Download AVANCE III NMR Hardware User Guide - Pascal-Man

Transcript
Cabinet and System Overview
•
•
operate the sample lift and spin
generate gradient pulses (where a GAB/2 unit is fitted).
For an overview of some of the BSMS signals see "Simplified Schematic of
Principal Signal Paths" on page 30.The BSMS hardware is comprehensively
described in various manuals available on the BASH CD and will not be dealt with
in any great detail in this manual with the exception of some recently developed
boards such as the new
•
•
•
ELCB which will be described in chapter 6.
GAB/2 which will be described in chapter 7.
SCB20 which will be described in chapter 8.
A summary of the various boards follows and they will be discussed in the order in
which they are located (from the left) as in Figure 2.3.
Deuterium Amplifier 2H-Tx: This houses a single 20W amplifier primarily designed
to facilitate experiments that require deuterium decoupling.Note that there is also
an AQS version with 80W power which is enough to observe deuterium although
the 90 degree pulses are relatively long.There is also a BLAXH2H external amplifier which delivers 150W on the deuterium channel (up to 600 MHz systems) and
250W on the deuterium channel (700-900 MHz systems).The BSMS version mentioned above has limited power (20W) due to the current limit of the BSMS power
supplies.
Sample and Helium Level Control Board SLCB: This board is responsible for all
aspects of raising, lowering and spinning the sample as well as monitoring the Helium level. The pneumatic module which controls a supply of compressed air to
drive the lift and spin is located at the rear of the BSMS/2.
Gradient Amplifier Board GAB/2: This board receives digital data from the G-Controller and converts it into its analogue equivalent in order to perform Z axis gradient spectroscopy. For further details see chapter 7.
Shim Current Board SCB/20.This generates the shim currents for the room temperature shims based on the operator adjustments at the BSMS Keyboard, Shuttle or from within TOPSPIN. Effectively the board is a very sophisticated DAC with
high accuracy and resolution.Figure 2.3. shows a BSMS/2 with two of these
boards fitted. The actual number of boards fitted (one or two) will depend on the
complexity of the shim system. The font panel connections where the currents are
accessed are labelled in Figure 2.3.. From here the cabling ports the currents to
the rear of the cabinet and then on to the actual shim system located within the
magnet bore. The SCB20 will be dealt with in chapter 8.
Extended Lock Control Boards ELCB: This new board replaces the old LCB and
CPU. In its role as CPU it controls the BSMS and in particular the lock. The main
development has been the Ethernet capability. It also incorporates 2 TTY connectors for BACS, VTU, etc.For more details see chapter 6.
Lock Receiver Board LRX: Like the Transmitter this board is spectrometer frequency dependent and receives the NMR signal generated by the deuterated solvent. The signal is transmitted from the probe via the 2H module of the HPPR or
internal preamp.
Lock Transmitter Board LTX: This board generates the deuterium excitation signal required to maintain the lock and as such is spectrometer frequency dependent. Figure 2.3. shows a LTX600 signifying that is fitted on a 600 MHz
spectrometer. The signal generated by this board is transmitted to the 2H module
of the HPPR or internal preamplifier and then to the probe.
User Guide Version 001
BRUKER BIOSPIN
23 (235)