Download Owner / Calibration Manual
Transcript
Model 8903B General Information lbble 1-2. Supplemental Information All parameters describe performance in automatic operation or with properly set manual controls. SOURCE Frequency Switching Speed: <3 ms (does not include HP-IB programming time). Sweep Mode: Logarithmic sweep with up to 500 points/decade or 255 Points total between entered start stop frequencies. Output Level Switching Speed: 20 ms (does not include HP-IB programming time. Typical Source to Analyzer Isolation (Option 001 only) System distortion and noise can be degraded when driving low impedance loads (with source output impedance in the 50R mode) due to coupling from the source output to the analyzer input. See Figure 1-7 for the plot of various load impedances vs frequency. MEASUREMENT SINAD Detection: True rms or rms calibrated average. Resolution: 0.01 dB for ratios >25 dB. For ratios <25 dB, the display is rounded to the nearest half dB to reduce digit flickering with noisy signals. (Full resolution is available by defeating this feature using Special Function 16.1 .) Analog Meter* active in SINAD only and for SINAD ratios 518 dB (or 524 dB using Special Function 7.1). Accuracy: 1 dB typical. Tuning: notch filter is tuned to analyzer source frequency. Time to Return First Measurement: 1.5s typical. Measurement Rate: 2.0 readingsls typical. SIG/NOISE Resolution: Same as SINAD. Measurement Rate: One readinglsecond for rms and average detectors (one reading/2 seconds for quasipeak detector). Detection: True rms or rms calibrated average or quasipeak. Time of Return First Measurement: ~ 2 . 5 stypical. Operation: The Audio Analyzer displays the ratio of the input voltages as the internal source is switched on and Off. DISTORTION 3 dB Measurement Bandwidth: 10 Hz to 500 kHz Detection: True rms or rms calibrated average. Displayed Resolution: 0.0001% (<0.1% distortion) 0.001% (0.1% to 3% distortion) 0.01OO/ (3% to 30% distortion) 0.1YO(>30% distortion) Time to Return First Measurement: 1.5s typical. Measurement Rate: 2 readings/s typical. ~~ AC LEVEL AC Converter: True rms responding detection for signals with (crest factor of 53), rms calibrated average detection, quasi-peak detection (Meets CClR 468-3 standard.) 3 dB Measurement Bandwidth: >500 kHz. Time to Return First Measurement: ~ 1 . 5 typical. s Measurement Rate: 2.5 readings/s for rms and average detectors. Quasi-peak Detector Accuracy: (20 Hz to 20 kHz) f6%. DC LEVEL: s Time to Return First Measurement: ~ 1 . 5 typical. Measurement Rate: 3 reading/s. FREQUENCY MEASUREMENT Measurement Rate: same as measurement mode selected. Counting Technique: reciprocal with 2 MHz time base. AUDIO FILTERS 400 Hz High-Pass Filter Rejection: >40 dB at 240 Hz; >65 dB at 60 Hz. REAR-PANEL INPUTS AND OUTPUTS Recorder Outputs: X Axis: 0 to 10 Vdc corresponding to the log of the oscillator frequency. Output Resistance: 1kR Y Axis: 0 to 10 Vdc corresponding to the displayed value and entered plot limits Output Resistance: 1 kR Pen Lift: TIL output Monitor Output: Output Impedance: 600R In ac level mode, provides scaled output of measured input signal In SINAD, distortion, and distortion level modes, provides scaled output of input signal with the fundamental removed. ~~ The meter has hysteresis. The trip point for the 24dB range is > 18 dB, and the trip pont for the 18-dB range is <17 dB. 1-22 rev.30DE C88