Download SigmaPace 1000 External Pacemaker Analyzer
Transcript
SigmaPace 1000 Operators Manual Command Syntax Commands can be sent in upper or lower case. All command strings must be terminated with a carriage return [CR] and/or a line feed [LF] in the format [command][CR][LF]. Spaces are ignored. A command shown with an equals sign (=) is a setup command that requires at least one parameter to be included with the command. Multiple parameters are separated by commas. All possible parameter choices are shown. Unless otherwise stated, the default parameters are listed first. Responses to Commands The instrument responds to all commands. Commands that do not return any other message or data return the confirmation, OK. After receiving a command, the instrument ignores all other incoming characters until it has responded to the one at hand. Remote Mode Analyzer Setup Commands The following are common commands associated with setting up remote operation of the Analyzer. VER.XX The firmware version level is returned in a three-digit format with a fixed decimal point. (Example: VER = 1.03) Beta version firmware is identified with an additional pair of alphanumeric characters (Example: VER= 1.03B1) IDENT The firmware returns the identity of the Analyzer name and firmware revision (Example: IDENT= ANALYZER 1.03) 5-4