5 Function modules from A - Z
11. Open the measuring chambers (see Fig. 86, 23 and 24) and pull off the
Measuring chamber output tube with a flat pair of tweezers.
12. Loosen the 5 screws on the measuring chamber cartridge and remove it.
NOTE:
Before replacing the electrode plugs clean their coupling piece
on the measuring chamber with cotton swab.
NOTE:
Do not use a sharp edged pair of tweezers to remove the tubing.
Components from the AVL OMNI  such as tubing, Waste
container, fill port etc. contain biological fluids after use,
resulting in possible risk for infections. Handle these
components carefully, and avoid contact with the skin.
Use rubber gloves.
The tubes may drip a little after being disconnected. Remove
excess fluids with a clean, absorbent cloth.
The measuring chamber tubing can be connected easily by disassembling the
electrode holder.
After reassembling a MC cartridge, please confirm the display on the screen
with "Continue" and change back to the "Ready" - screen.
5-72
Service Manual, AVL OMNI, Rev. 9.0, May 2000