5 Function modules from A - Z 11. Open the measuring chambers (see Fig. 86, 23 and 24) and pull off the Measuring chamber output tube with a flat pair of tweezers. 12. Loosen the 5 screws on the measuring chamber cartridge and remove it. NOTE: Before replacing the electrode plugs clean their coupling piece on the measuring chamber with cotton swab. NOTE: Do not use a sharp edged pair of tweezers to remove the tubing. Components from the AVL OMNI such as tubing, Waste container, fill port etc. contain biological fluids after use, resulting in possible risk for infections. Handle these components carefully, and avoid contact with the skin. Use rubber gloves. The tubes may drip a little after being disconnected. Remove excess fluids with a clean, absorbent cloth. The measuring chamber tubing can be connected easily by disassembling the electrode holder. After reassembling a MC cartridge, please confirm the display on the screen with "Continue" and change back to the "Ready" - screen. 5-72 Service Manual, AVL OMNI, Rev. 9.0, May 2000