EVENT
The wheelchair is not
locked in position in
Permolock.
POSSIBLE CAUSE
You have forgotten to press
the locking device control
button before moving the
wheelchair over the locking
plate.
It is more than 40 seconds
since the locking device
control button was pressed.
REMEDY
Reverse the wheelchair off the
locking plate. Press the locking
device control button so the
locking device extends, before
moving in over the locking
plate again.
The wheelchair is not
locked in position in
Permolock.
The manual release handle
is in open position.
Move the manual release
handle to locked position.
The locking mechanism’s
setting tool is broken.
Contact service
The wheelchair is not
adapted to Permolock C
when driven in over the
lock.
Air pressure in wheelchair
tires is too low.
Adjust tire pressure, see
wheelchair user manual.
Permolock’s rubber stops
are wrongly adjusted.
Adjust the rubber stops as
instructed on page 25.
The wheelchair’s shock
absorbers are wrongly
adjusted.
Adjust the shock absorbers,
see wheelchair service
manual.