Download - Frank`s Hospital Workshop
Transcript
E03 T2 TEMPERATURE SENSOR ERROR Detection T2 temperature exceeding measurable limit Detection Audible process alarm Rinse wait ○ (7 times) Continuance Rinse ↑ ↑ Rinse end ↑ ↑ Air purge ↑ ↑ Dialysis ○ (Continuous) ↑ ↑ Dialysate line stop Preparation ↑ ↑ Prep. end ↑ ↑ Dialysis ↑ ↑ Dialysate line function wait BP/SP stop CLV function Reset ― ― ○ Recovery Rese After Ver 3, measurable extent of upper limit while Press [RESET] SW. Note t rinse isn’t detectable. Condition or cause Check and action Check: Check whether indications of all temperature sensors and other sensors such as dialysate pressure sensor are incorrect. Indication error of temperature and Check stable output of voltage of ≧4.9V (nominal: 5Vref) at test other analog system point on main board. Action: Replace main board with new one. Check: Check no incorrect wiring, loosening, corrosion and/or disconnection at TM1 T2 terminal of main board. Incorrect [T2] terminal connection Action: If loosened, tighten securely. If incorrectly wired, correct. If corroded or disconnected, replace [T2] with new one Check: Remove TM1 T2 terminal from main board and measure resistance between Red and White to find short circuit, Internal disconnection or short circuit disconnection and fluctuation. of [T2] Action: Replace [T2] with new one. Action: Calibrate temperatures indicated by [T1], [T2] and [T3] circuits Incorrect indicated temperature of main board using temperature calibrator. calibration of main board If the calibration is impossible, replace main board with new one.