(3) Filter sensor
Open circuit (E05): When the temperature in filter sensor is equal or higher than 130㷄, E05
and chamber temperature are displayed alternately, the buzzer beeps
intermittently and remote alarm contact activates.
High side compressor is forcibly turned off.
Press BUZZER key to stop the buzzer beeping.
Short circuit (E06): When the temperature in filter sensor is equal or lower than -60㷄, E06
and chamber temperature are displayed alternately, the buzzer beeps
intermittently and remote alarm contact activates.
Press BUZZER key to stop the buzzer beeping.
(4) AT sensor
Open circuit (E07): When the temperature in AT sensor is equal or higher than 60㷄, E07
and chamber temperature are displayed alternately, the buzzer beeps
intermittently and remote alarm contact activates.
Unit should have warming-up operation when low side compressor is
turned on.
Press BUZZER key to stop the buzzer beeping.
Short circuit(E08): When the temperature in AT sensor is equal or lower than -60㷄, E08
and chamber temperature are displayed alternately, the buzzer beeps
intermittently and remote alarm contact activates.
Unit should have warming-up operation when low side compressor is
turned on.
Press BUZZER key to stop the buzzer beeping.
(5) Battery SW is off If you press ALARM TEST key when battery switch is off position or
position (E09):
battery is unconnected, E09 is displayed.
(6)Compressor
abnormal
temperature (E10):
When the temperature in filter sensor is equal or higher than 60㷄, E10
and chamber temperature are displayed alternately and high side
compressor is forcibly turned off to inform compressor temperature is
abnormal or fan motor locked.
Press BUZZER key to stop the buzzer beeping.
When the temperature in filter sensor – AT is equal or lower than 10㷄,
compressor is turned on.
(7) Error code priority
No.1:
No.2:
No.3:
No.4:
No.5:
6.
Alarms
High temp. alarm
:
Cascade sensor (E03, E04) … High side compressor is forcibly turned on
Filter sensor (E05, E06) … Compressor protection is uncontrollable
Abnormal compressor temp.(E10) … Compressor temporary turns off
Temp. sensor (E01, E02) … Compressor keeps forcibly running
AT sensor (E07, E08) … Warming-up operation in any AT
When chamber temperature is equal or higher than set temperature +
high temp. alarm set temperature +1.0, ALARM lamp and LED display
blinks, intermittent buzzer beeps with approx. 10 minutes of delay and
remote alarm activates.
When chamber temperature is equal or lower than set temperature,
ALARM lamp and LED display go off, buzzer stops beeping and remote
alarm turns off.
If you press BUZZER key, the buzzer stops beeping instead remote
alarm does not turn off.
17