Download Service Manual - Panasonic Biomedical
Transcript
(3) Filter sensor Open circuit (E05): When the temperature in filter sensor is equal or higher than 130㷄, E05 and chamber temperature are displayed alternately, the buzzer beeps intermittently and remote alarm contact activates. High side compressor is forcibly turned off. Press BUZZER key to stop the buzzer beeping. Short circuit (E06): When the temperature in filter sensor is equal or lower than -60㷄, E06 and chamber temperature are displayed alternately, the buzzer beeps intermittently and remote alarm contact activates. Press BUZZER key to stop the buzzer beeping. (4) AT sensor Open circuit (E07): When the temperature in AT sensor is equal or higher than 60㷄, E07 and chamber temperature are displayed alternately, the buzzer beeps intermittently and remote alarm contact activates. Unit should have warming-up operation when low side compressor is turned on. Press BUZZER key to stop the buzzer beeping. Short circuit(E08): When the temperature in AT sensor is equal or lower than -60㷄, E08 and chamber temperature are displayed alternately, the buzzer beeps intermittently and remote alarm contact activates. Unit should have warming-up operation when low side compressor is turned on. Press BUZZER key to stop the buzzer beeping. (5) Battery SW is off If you press ALARM TEST key when battery switch is off position or position (E09): battery is unconnected, E09 is displayed. (6)Compressor abnormal temperature (E10): When the temperature in filter sensor is equal or higher than 60㷄, E10 and chamber temperature are displayed alternately and high side compressor is forcibly turned off to inform compressor temperature is abnormal or fan motor locked. Press BUZZER key to stop the buzzer beeping. When the temperature in filter sensor – AT is equal or lower than 10㷄, compressor is turned on. (7) Error code priority No.1: No.2: No.3: No.4: No.5: 6. Alarms High temp. alarm : Cascade sensor (E03, E04) … High side compressor is forcibly turned on Filter sensor (E05, E06) … Compressor protection is uncontrollable Abnormal compressor temp.(E10) … Compressor temporary turns off Temp. sensor (E01, E02) … Compressor keeps forcibly running AT sensor (E07, E08) … Warming-up operation in any AT When chamber temperature is equal or higher than set temperature + high temp. alarm set temperature +1.0, ALARM lamp and LED display blinks, intermittent buzzer beeps with approx. 10 minutes of delay and remote alarm activates. When chamber temperature is equal or lower than set temperature, ALARM lamp and LED display go off, buzzer stops beeping and remote alarm turns off. If you press BUZZER key, the buzzer stops beeping instead remote alarm does not turn off. 17