F6 Fetal & maternal Monitor User Manual
current measurement and insulation testing. The recommended testing interval is once per
year.
A The protective categories against electric shock of the patient connections are:
1) Ultrasound (FHR1, FHR2)
2) External TOCO
3) Fetal Movement Mark (FM)
4) Fetal Stimulator (FS)
This symbol indicates that the electric shock defend grade of this instrument is Type B.
1) IUP
This symbol indicates that the electric shock defend grade of this instrument is Type BF.
1) DECG
This symbol indicates that the electric shock defend grade of this instrument is Type CF.
The monitor described in this user manual is not protected against:
a) The effects of defibrillator shocks
b) The effects of defibrillator discharge
c) The effects of high frequency currents
d) The interference of electrosurgery equipment
1.3 Ultrasound Safety Guide
A Fetal Use
The monitor is designed for continuous fetal heart rate monitoring during pregnancy and labor.
Clinical interpretation of fetal heart rate traces can diagnose fetal and/or maternal problems and
complications.
A Instructions for Use in Minimizing Patient Exposure
The acoustic output of the monitor is internally controlled and can not be varied by the operator
in the course of the examination. The duration of exposure is, however, fully under the control of
the operator. Mastery of the examination techniques described in the User Manual will facilitate
obtaining the maximum amount of diagnostic information with the minimum amount of exposure.
The exercising of clinical judgment in the monitoring of low risk patients will avoid unnecessary
insonation.
1.4 Safety Precautions
WARNING and CAUTION messages must be observed. To avoid the possibility of injury,
observe the following precautions during the operation of the instrument.
-2-