Commands Listed in Alphabetical Order
Examples
SAVE:SETUP 5 saves the current oscilloscope setup in memory location 5.
SAVE:SETUP "TEK00000.SET" saves the current oscilloscope setup in the file
TEK00000.SET in the current working directory.
SAVe:WAVEform (No Query Form)
This command saves a specified waveform or all displayed waveforms (excluding
serial bus waveforms). Only individual analog waveforms (CH<x>, MATH and
REF<x>) can be saved to reference memory locations.
You can save all displayed waveforms, excluding serial bus waveforms, to a
single CSV file when the SAVE:WAVEFORM:FILEFORMAT is set to SPREADSHEET.
You can save all displayed waveforms, excluding serial bus waveforms and digital
channels (MSO models) to consecutive ISF (internal save format) files when the
SAVE:WAVEFORM:FILEFORMAT is set to INTERNAL.
Group
Save and Recall
Syntax
SAVe:WAVEform [<wfm>,{REF<x>}] | [<wfm>, <QString>] | [ALL,
<QString>]
Related Commands
Arguments
RECAll:WAVEform, SAVe:WAVEform:FILEFormat
<wfm>, <REF<x>> saves the specified waveform to the specified reference
memory location. <wfm> can be any live analog channel (where <x> is the
channel number), the MATH1 waveform, or another reference waveform (such
as REF1).
<wfm>, <QString> saves the specified waveform to the file specified in the
quoted string argument. Any live channel (such as CH1), the MATH1 waveform,
any reference waveform, and, for MSO models, digital waveforms (such as
D0) can be saved to a file. Digital waveforms can only be saved when the
SAVE:WAVEFORM:FILEFORMAT is set to SPREADSHEET.
ALL, <QString> saves all displayed waveforms, excluding serial bus
waveforms, to a single CSV file specified by the quoted string argument when the
SAVE:WAVEFORM:FILEFORMAT is set to SPREADSHEET, or saves all displayed
waveforms, excluding serial bus waveforms and digital channel waveforms (MSO
models), to individual ISF (internal save format) files with a file name prefix
specified by the argument with an underscore (_) and the waveform ID (such as
CH1, REF1, MATH) appended to the file name(s).
2-232
Tektronix 4000 Series Programmer Manual
Related documents
TRIGger:A:BUS:B
CH? (Query Only)
Tektronix DPO4000 User's Manual
Oscilloscope Programmer`s Reference
Connect-Tek USB-2 Specifications
Programmer Manual TDS 340A, TDS 360 & TDS 380 Digital Real
MEASUrement:MEAS? (Query Only)
SEARCH:SEARCH
Models 2603, 2621, and 2635 IPLink Series High
Tektronix scope programmer`s manual
Pocket Tracker Kit Illustrated Assembly Instructions
Appendix A Specifications and Quick Starts