TXgard-IS+
Commissioning
Use  and  buttons to adjust gas reading to calibration
gas value.
Gas value is usually measured in ppm or as %vol (ppb and
%LEL are also available, as appropriate).
Tip
See "Table 4.1" on page 21 for some gases and their typical
range. Please note that detectors can be provided in different
ranges if required.
Contact Crowcon for list of all gases detected.
8 Select the  button to set the calibration value.
The instrument will display the following message:
Calibrating..
TXgard-IS+
When complete the instrument will report a scrolling text
message: ‘Calibration successful! Remove gas
from the instrument ––’.
9 Remove the gas from the detector and exit the menu
system.
The instrument is now calibrated.
If the control equipment display requires adjustment, consult the operating
manual for the control display equipment.
20