Download Crowcon Txgard-IS+ Fixed Gas Detector
Transcript
TXgard-IS+ Commissioning Use and buttons to adjust gas reading to calibration gas value. Gas value is usually measured in ppm or as %vol (ppb and %LEL are also available, as appropriate). Tip See "Table 4.1" on page 21 for some gases and their typical range. Please note that detectors can be provided in different ranges if required. Contact Crowcon for list of all gases detected. 8 Select the button to set the calibration value. The instrument will display the following message: Calibrating.. TXgard-IS+ When complete the instrument will report a scrolling text message: ‘Calibration successful! Remove gas from the instrument ––’. 9 Remove the gas from the detector and exit the menu system. The instrument is now calibrated. If the control equipment display requires adjustment, consult the operating manual for the control display equipment. 20 Issue 7