Chapter 2
SYSTEM SETUP
Valve
Fluid
Inlet/Outlet
Port
Safety
Disk
Pump
Cylinder
Figure 2-19
Pressure
Transducer
Figure 2-20
2. Install the tubing from the center
port of the valve into the fluid
inlet/outlet port with the tube that
is attached to the valve. The valve
will extend out from the end of the
cylinder barrel. Refer to Figure 220. Repeat for each pump cylinder
in your system.
Valve
3. Install the tubing from the
pressure transducer into the lower
left hole at the end of the cylinder
Pressure Transducers
barrel. See Figure 2-21. For
ambient temperature systems,
Figure 2-21
mount the transducer to the side of
the cylinder barrel. For high temperature systems, the transducer must be placed outside
the oven if the temperature will exceed 80 degrees C.
4. The third hole at the end of the cylinder barrel has a safety rupture disk
installed. There will be a tag indicating the
specifications of the safety rupture disk. See Section
5.3.4 for more information on this important safety
feature.
Air Tube
2.12.2 Q6000-20K Pump Systems
For the Q6000-20K pump systems the pump cylinders
are designed for a maximum pressure of 20,000 psi and
have only one port at the end of the cylinder barrel. The
tubing from this port attaches to a cross (a 4-holed tee).
From the cross, attach tubing to connect to (1) the valve,
(2) the pressure transducer, and (3) the safety rupture disk
assembly.
Figure 2-22
2-15