Download LOBO System Operation Manual

Transcript
System Operation Manual
LOBO
SAT-DN-00434, Rev. B
2011-11-24
Page 79 of 85
11.9 Seabird SBE-37SIP Maintenance
Section 5 of the SBE-37SIP user manual contains detailed maintenance information. Key points:








WARNING! Do not run the pump dry!
Rinse with fresh water after each use
Avoid direct electrical connection of the SBE-37SIP body to other metal surfaces
Lubricate connectors prior to connecting (DC-4 or equivalent)
Refer to the manual for detailed conductivity cell maintenance instructions
Periodically inspect the pressure port to remove any particles or debris
Replace anti-foulant as required – please contact Sea-Bird for guidelines
Return the sensor to Sea-Bird periodically (annually) for recalibration
11.10 Aanderaa Oxygen Optode Maintenance
Chapter 5 of the optode user manual (RD7) contains detailed maintenance information. Key
points:





The optode should be cleaned at regular intervals (1 month to 1 year) depending on the
required accuracy and fouling conditions at the deployment site. Use clean fresh water for
cleaning; use a wet cloth for the sensing foil and a brush for the housing. Vinegar can be
used on calcareous organisms.
A copper plate immediately in front of the optical window can be used to extend intervals
between cleaning, as it prevents fouling from occurring directly on the window.
A copper mesh (such as a cleaning pad) could also work, although it may trap sediment. Be
careful to avoid dissimilar metal contact.
Recalibrate the sensors annually.
Replace sensing foil and recalibrate if it becomes damaged.
11.11 Solar Panel Maintenance
Occasional or as-required cleaning of the solar panel surface with soapy water (such as
dishwashing liquid) and rinsing with fresh water is the major maintenance requirement. The
cleaner the surface, the better the light transmission, allowing the panel to operate at or close to
rated capacity. Bird droppings are notoriously bad for reducing light transmission! Avoid
scratching the glass surface.
In addition, rinse the frame with fresh water to remove salt crystals – this will help reduce
corrosion.
11.12 Battery Box Maintenance
Little maintenance is required for the battery assembly. Key points:







Remove fuses prior to storage. This will help prolong battery charge, as it will remove power
from the CLC, which draws a small but constant current from the battery.
Rinse the outside of the box periodically to remove salt deposits – this should help reduce
corrosion.
Lubricate male connector pins lightly with DC-111 or equivalent prior to mating with the
battery box.
Store with dummy connectors in place
Lightly lubricate male connector pins with Dow Corning DC-111 lubricant or equivalent
Any debris (sand, mud) in the female contact sockets should be removed with fresh water to
prevent damage to the contact and o-ring. Remove the fuses first so that the connector pins
are not live!
Do not disconnect cables by pulling on the cables. Pull straight back on the connector.
Copyright  2011 Satlantic Inc. All rights reserved.