Download 1260 Infinity Multi-Detector Suite User Manual
Transcript
3 Installation Other Connections Auxiliary I/O Port CAUTION The Pump stop facility must be employed if the instrument is to be left unattended. ➔ Connect the auxiliary cable ( Agilent 1260 Infinity MDS Auxiliary I/O cable (G7800-63003)) between MDS auxiliary port and remote port on 1200 Series LC System. The MDS is fitted with an auxiliary 15 pin I/O connection for connecting the supplied auxiliary I/O cable. The auxiliary I/O cable provides the following: • A ‘Stop Pump Request’ allowing a pump to be stopped if any error occurs on the MDS. • Receive a sample injection trigger for synchronized data acquisition. Table 9 Inputs Outputs MDS I/O Connections I/O Description Pin Number Wire ID Sample Injection Trigger Input 5 & Ground 7 & GND Spare TTL Input 6 8 Pump Stop Contact Closure - Normally Open 3 & 10 1&2 Sample Injection Trigger Output - Normally Open 4 & 11 5&6 Pump Stop - TTL Active Low 2 & Ground 4 & GND Pump Stop - TTL Active High 9 & Ground 3 & GND Ground (to case) 1, 7, 13 & 14 GND Serial RS232 Port The MDS is fitted with two serial RS232 ports. These two ports are reserved for future use. 56 Agilent MDS User Manual