Model LTC-21 Cryogenic Temperature Controller
resolution. All loop variables, including the set-point, are also in 32-bit floating
point.
• High
frequency damping is performed on the differentiator term so that effective use
of this term can be allowed without instabilities caused by high frequency noise.
• Safety monitoring is
implemented so that an open or shorted connection to the input
sensor will be detected and will cause the control loop to be terminated and the
heater to be turned off.
Heater Autotune Function.
The heater control has an Autotune function wherein the instrument will 'learn' the
thermal characteristics of the system under control; then, generate appropriate PID
coefficients required to efficiently control it.
Analog Output
A low power analog output channel provided. This output can be used as a linear output
or as a second heater for temperature control of low-power systems.
Output voltage is ±12 V with a maximum current of 100mA. This is a voltage output and
over-current protection is NOT provided. Therefore, this output can be destroyed if
connected to a low resistance load that will require an excess of 100mA.
Connection to the analog output channel is made via a 4-pin connector on the back
panel. The connector housing provides a shield ground.
CAUTION
When using
output, it is
power setting
can
the analog output as a heater
important to remember that at 0
the output is not at 0 volts but
be as much as ± 100mV.
The analog output may also be used to drive a user provided
external power supply, resulting in a heater output ranging
from 30W to kWs.
Control Channel Configuration
Each output (Heater or Analog) may be assigned to either input channel independently.
Therefore, all combinations of monitoring one channel and controlling another are
allowed.
If the Heater output and the Analog output are configured for temperature control, the LTC21 will maintain two completely separate control loops.