Download Manual - Test Equipment Depot
Transcript
10-4 Analog Output Temperature The analog output voltage for temperature measurements depends on thermocouple type and the selected units (°C, °F, or K). The 1.2V analog output is scaled to the maximum positive temperature reading. For example, the measurement range for the Type J thermocouple is -200°C to +760°C. For a 760°C reading, the analog output voltage will be 1.2V, and for a -200°C reading, the analog output voltage will be –0.316V. The measurement ranges (in °C) for the various thermocouple types are listed in the specifications (see Appendix A). With Gain set to 1 and Offset set to 0, analog output voltage for temperature measurements is calculated as follows: Analog Output = 1.2 × Rdg/Rng* * Rng is a magnitude. Therefore, it is always a positive value. Example Calculations: Type J, 100°C Reading Type J, –100°C Reading Analog Output = 1.2 × 100/760 Analog Output = 158mV = 1.2 × (–100/760) = –158mV When using Gain and Offset, the calculation is expanded as follows: Analog Output = (Gain × 1.2 × Rdg/Rng) – Offset Ratio Analog Output can be used with Ratio. When enabled, the analog output voltage is scaled to a Ratio value of 1. That is, the analog output will be 1V for a Ratio result of 1. If, for example, the Ratio is 0.4, analog output voltage will be 0.4V. Gain, Offset, and Analog Output Rel can also be used. However, the maximum analog output is ±1.2V. Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com