9 Troubleshooting and corrective actions
9.2 Problems during a run
Elution problems
Find the possible cause and action for a specific problem in the table below.
No satisfactory elution
from...
Possible cause and action
Affinity column
If the protein did not bind to the column, it will be
found in the flowthrough that was collected in one
of F3–F6.
• Verify that the tagged target protein is present in
the start sample.
• Check that the sample has been applied onto the
affinity column.
• Check that appropriate wash and elution buffers
were applied.
• HisTrap HP and HiTrap Chelating HP:
- If the target protein elutes during Wash 2, decrease the imidazole concentration of wash 2
solution.
- If the absorbance curve does not return completely to the base line after peak elution, the
quality of the imidazole might be poor. Change
to imidazole with higher quality, or increase
the Peak_Max factor to be able to detect peak
end.
• GSTrap HP and FF:
- Many proteins bind better to GSTrap HP and
FF at room temperature than in cold room
whereas other protein shows the opposite.
- Adding 1 mM DTT or other reducing agent to
the buffer might increase the binding capacity.
• Tag cleavage:
- High glycerol concentrations in the buffer might
decrease the activity of some proteases.
- For operation and optimization hints see 4
Methodology overview on page 53 and 6.5.8
Performing automatic affinity tag removal on
page 175.