R&S®SGT100A
Remote Control Commands
Fast Speed Commands
The instrument setting must first be stored in an internal memory with the same number using the common command *SAV.
Setting parameters:
<savrcl_state_nr>
Corresponds to the specific <number> defined with the *SAV
command, e.g. *SAV 4.
<file_name>
String parameter to specify the file name with extension
*.savrcltxt.
Usage:
Event
11.11 Fast Speed Commands
This section describes special commands that allow a fast frequency and level setting.
:FFASt.......................................................................................................................... 284
:PFASt.......................................................................................................................... 284
:FFASt <Freq>
Special command to set the RF output frequency with minimum latency. No unit (e.g.
Hz) allowed.
Bypasses the status system so command *OPC? cannot be appended.
Parameters:
<Freq>
float
Example:
FFASt 12750000000
:PFASt <Pow>
Special command to set the RF output level with minimum latency. No unit (e.g. dBm)
allowed.
Bypasses the status system so command *OPC? cannot be appended.
Parameters:
<Pow>
float
Example:
:PFASt -20
11.12 OUTPut Subsystem
:OUTPut:AFIXed:RANGe:LOWer?................................................................................... 285
:OUTPut:AFIXed:RANGe:UPPer?....................................................................................285
:OUTPut:AMODe........................................................................................................... 285
:OUTPut[:STATe]...........................................................................................................285
:OUTPut[:STATe]:PON................................................................................................... 285
User Manual 1176.8674.02 ─ 06
284