Download 1 - ECD Veterinaria
Transcript
C hapter 9 C alibration Test 1.Calibration Test Procedures Select Test Distilled Water Test Blank Test Calibration Test Next Calibration Test 1.The Rinsing process is not mandatory process, in order to avoid crossed pollution, Rinse with distilled water after the completion of each test. 2.Please follow the blank test of the Standard liquid when the parameter setting is Standard Blank Test. Figure 9-1 This instrument provides two calibration tests: linear and non-linear. E.g. ALT calibration Test Procedures: ¾ 1) Turn on the instrument to enter the main menu. Click 【Test】to the【Select Animal】 interface ,select a animal and click 【OK】,then enter the 【Select Test】. Select ALT and click Enter (or double click ALT) to【Test】interface. Click 【Edit】to the 【Test Parameter Setting】interface. Please set the parameters if the parameters existed could not comply with the operation manual of the Reagent. ¾ 2) The system will adjust the Lamp and Temperature automatically. The real time temperature will be displayed on the top right corner of the interface. The operation QRWLFH³3OHDVHaspirate GLVWLOOHGZDWHU´ZLOOGLVSOD\DIWHUWKHDGMXVWPHQW 55