Download Chapter 15 Accuracy and Repeatability
Transcript
Chapter 15 Accuracy and Repeatability Examples — Determining System Accuracy and Repeatability Dimensions: see figure below Maximum distance measured (L): 1.0 m Deadpath distance (D): 0.1 m Cosine Error: 0.05 ppm (Agilent 10716A aligned according to procedure in this manual) Nonlinearity: ±1.0 nm (Agilent 10716A) Abbé error: none (assume zero offset) Measurement resolution: ±2.5 nanometers (Agilent 10716A) ENVIRONMENT: Temperature: 20 °C ±0.5° (temperature controlled environment) Pressure: 760 mm Hg ±25 mm Hg (possible storm fronts during measurement, pressure not controlled) Humidity: 50% ±10% (humidity controlled environment) LASER SYSTEM CONFIGURATION ON CMM Agilent 10716A High resolution Interferometer Agilent 10780C Receiver Agilent 10716A High Resolution Interferometer Agilent 10707A Beam Bender Agilent 10724A Plane Mirror Reflector Agilent 10701A 50% Beam Splitter Agilent 10724A Plane Mirror Reflector Agilent 10700A 33% Beam Splitter Agilent 10780C Receiver Agilent 10707A Beam Bender Agilent 10780C Receiver Z X Agilent 10717A Wavelength Tracker Y Agilent 10707A Beam Bender Agilent 10707A Beam Bender Agilent 10724A Plane Mirror Reflector Agilent 5517B Laser Head Agilent 10716A High Resolution Agilent 10780C Agilent 10700A Interferometer Receiver 33% Beam Splitter Figure 15-9. Laser system configuration for a precision Coordinate Measuring Machine (CMM) 15-28 User’s Manual