enabling innovative electronics Safety/operational considerations The Ossila Multiplexer Measurement Unit was designed to be used with Drain (Channel B) and Gate (Channel A) currents and voltages below 100 mA and ±100 V, respectively. The Ossila SuperFACT Control Software automatically limits both output currents and driving voltages to comply with these limits. Due to this reason, and in order to avoid any signal degradation, there is no safety fuse in the signal paths. If not using the Ossila SuperFACT Control Software, ensure the Drain (Channel B) and Gate (Channel A) currents and voltages signals are always limited to 100 mA and ±100 V, respectively. These limits should be hard-coded in the control software so that no user can unintentionally supply/read voltages/currents whose values are outside the operational range of the multiplexer. If using alternative software, is the software engineer/programmer is responsible for testing whether the application is working correctly and the SMU is operating within the limit described above It is also strongly recommended to include error-handling routine to make sure the application will attempt to disarm the channel whenever an error occurs. The multiplexer is equipped with an interlock on the top panel (figure 7). This is a safety feature that Interlock Figure 7. Interlock on the top panel. disconnects the probes when the lid is open. Note No measurement can be taken unless the lid is properly closed and the interlock is disengaged. Ossila Ltd Copyright © 2009-2015 20