MacTempas User Manual
tion; that is, underfocus is negative.
Aperture Radius
0.67
The value for the radius of the objective aperture should correspond to the radius in reciprocal Ångstrøm units, as measured
from a diffraction pattern exposed with the aperture superimposed. We will enter 0.67 to represent a typical value.
Center of the Objective Aperture
0,0
In order to simulate dark-field images, MacTempas provides for
an objective aperture displaced from the center of the diffraction pattern. As for the Laue circle center, the aperture center is
defined in units of h and k. We leave the default values of 0,0.
Center of the Optic Axis
0,0
To provide for microscope misalignment, or for conditions of
tilted-beam imaging, the coordinates of the diffraction pattern at
which the optic axis lies can be specified in the same manner as
the center of the aperture. Again, we use default values of 0,0.
Verifying the
Input
Running the Calculation
After the response to the last data-entry prompt, MacTempas
draws the windows it uses. To re-display the input information
click on the "Main Parameters" in the Parameters menu. At
this stage any desired changes can be made by using the mouse
to move the cursor to the desired parameter, and making the
change.
When all the data in the top field are satisfactory, we go to
"Atomic Basis" in the parameters menu to check that all atom
parameters have been entered correctly. At this stage it is also
worthwhile getting MacTempas to display a model of the structure by.going to the Commands menu and clicking on "Draw
the Unit Cell".
When we are satisfied that all data are correct, we run the simu-
Ch. 8 Sample Calculation - p.125