Appendix 7.8. for additional information regarding 384-well plates and small volume reactions. Note: Post-reaction cleanup is most easily and inexpensively achieved by ethanol precipitation using the sodium acetate/EDTA buffer provided with the kit. This buffer contains 1.5 M sodium acetate (pH > 8.0) and 250 mM EDTA. EDTA dramatically reduces the amount of unincorporated terminators that precipitate with the DNA, virtually eliminating dye blobs that distort sequence data near the primer and in adjacent lanes. Note: Do not combine the sodium acetate/EDTA buffer and ethanol prior to addition to the sequencing reaction. EDTA has very low solubility in ethanol and will precipitate. 7. Centrifuge the tubes at room temperature in a microcentrifuge for 15 minutes at ~ 12 000 rpm. Centrifuge 96-well or 384-well plates for at least 30 minutes at 2 500 x g or greater. 8. Remove the supernatant by aspiration from each microcentrifuge tube. For plates, a brief inverted spin (< 1 minute at 300 x g) is sufficient for supernatant removal. Remove as much liquid as possible at this step to prevent dye blobs. 9. Wash the DNA pellets with 70% ethanol. Use as large a volume of 70% ethanol as the tube or well can safely accommodate. Centrifuge briefly. Note: Scientists at GE Healthcare routinely use 250–500 µl for 0.5 ml microcentrifuge tubes, 100 µl for 96-well plates, and 50 µl for 384well plates. 10. Remove the supernatants by aspiration or by an inverted spin. Air-dry (preferably) or vacuum-dry (in a vacuum centrifuge) the pellets for 2–5 minutes. Do not overdry. 14