Download Insulation / Continuity MI 3121 Instruction manual
Transcript
MI 3121 Smartec Insulation / Continuity Continuity 5.2.3 Compensation of test leads resistance This chapter describes how to compensate for test leads resistance in both continuity functions (RLOW and CONT). Compensation is required to eliminate the influence of test leads resistance and the internal resistances of the instrument on the measured resistance. The lead compensation is therefore a very important feature to obtain correct result. Once compensation has been performed, the compensation symbol ( ) appears on the screen. Each of RLOW and CONT has each own compensation. Circuits for compensating the resistance of test leads Figure 5.9: Shorted test leads Compensation of test leads resistance procedure Select the any continuity function using the function switch. Connect test cable to the top of the instrument and short the test leads together (see figure 5.9). Press TEST to perform resistance measurement. Press the CAL key to compensate leads resistance. Note: The limit value for lead compensation is 5 Ω. 26