Download Operator`s Manual PDF

Transcript
Ellipse Plus Range –Operator’s Manual – Your Ellipse System
1.11 Telangiectasias and Reticular Vessels (Nd:YAG
Treatment)
Initial Notes:
Ellipse MultiFlex+ users with an Nd:YAG applicator are able to treat leg
telangiectasias and reticular vessels with a diameter between 0.1mm and
3mm. Treatment of vessels with a larger diameter is not allowed.
An extremely thin layer of optical coupling gel is required; this must not be
allowed to dry out during treatment, so apply gel only to a small area at a
time.
On legs: treatments using intense pulsed light are not recommended – the
preferred treatment is use of Nd:YAG.
Caution! the operator should test the air flow before shooting and allow
sufficient time for the air to cool the skin.
Target Chromophore: Hemoglobins
The target chromophores are hemoglobin and oxyhemoglobin in the blood
inside the vessels. The light absorbed by the blood is transformed into heat,
and then conducted to the vessel wall. The protein in the vessel wall will be
destroyed if it reaches 70°C for more than 1ms. As a result the blood will
coagulate and the destroyed vessel wall will gradually disappear.
Wavelength
The Nd:YAG applicator produces light with a wavelength of 1064nm.
Pulse Duration
Pulse duration is dependent on vessel size.
The thermal relaxation time of the vessel depends on its diameter. Thicker
vessels have a longer thermal relaxation time than thinner vessels. The
pulse duration must be close to the relaxation time of the treated vessel.
Your Ellipse system automatically selects appropriate pulse duration, when
the operator enters the thickness of the vessels.
Guide to Vessel Size
A vein gauge is supplied with the Nd:YAG applicator. This shows how
vessels of different diameters appear at the skin surface. Placing the gauge
next to the vessel allows the user to make a direct measurement of the
vessel size, and this information should be entered as a treatment
parameter onto the Ellipse system.
1MAN8439–A00-USA
Page 1-21