3
Installation
Installing the Solvent Delivery Module
Table 6
J2 terminal strip
Name
Connection
INJECT
Contact closure input from an external device which cancels a programmed
Inject Wait or Hold.
D GND
Digital ground for Inject signal1.
MARK
Contact closure input to perform an Event Mark (a 10 % vertical trace on
the chromatogram) from an external device.
D GND
Digital ground for Mark signal1.
1
Do not connect to Chassis Gnd
Setting the Pump ID and Pump Head Size When Using HPLC Control
Software
This is done on the pumps whether you are using an HPLC control software or
using one of the pumps as a master controller.
To set the pump ID when all pumps are slaves:
1 Turn the pump on.
2 Press SETUP and then press ID.
3 Set the ID for the Agilent 218 Pump either by entering a number between 0
and 63 , or pressing the UP ARROW or DOWN ARROW key to scroll through a
preset list of choices. Available choices are: 0 – 63 , MC (master controller)
or – – (no ID).
To set the pump head size when all pumps are slaves:
1 Turn the pump on.
2 Press SETUP and then press HdSz.
3 Use the UP ARROW or DOWN ARROW key to select between a preset list of
choices. Choices are: 5 , 10 , 25 , 50 , 100 , and 200 mL/min, 10P, 25P, 50P,
100P. The P designation stands for PEEK. The compressibility compensation
for PEEK heads is different than for stainless steel or titanium heads.
Purification Solution System User Guide
67