Download Operating Instructions NewClassic Balances ME
Transcript
Printout: ------ Formulation ----21.Jan 2012 12:56 Balance Type ME4002 SNR 1234567890 -----------------------1 T 10.33 g 1 N 8.85 g 2 N 9.23 g 2 T 10.84 g 3 N 7.43 g . . n 8 T Total 452.76 g G Total 546.79 g N Total 94.03 g ----------------------Function "FILL UP" This function allows you to add an additional component weight to the total weight of all compo nents to reach a desired target weight (Fill up). Starting the fill up function. FILL UP – Activate or deactivate function "FILL UP" by pressing « » (toggle). Filling up with an additional component weight: 94.03 Net C 100.00 Net F9 0.00 Net The last total of the component weights is displayed. 1 Add component weight until the desired target weight is reached. 2 Press « » to confirm. The display briefly shows the next component count marked with "F ", the current weight is stored as sample and the component weight is printed. The display is set back to zero. Filling up further additional component weights: The same procedure, beginning with starting up the "FILL UP" function. 54 Applications