Ion concentration MultiLab 4010-2 15. When calibrating without temperature sensor: Measure the temperature of the standard solution using a thermometer. Use <F4>/[ ⇅ ] to select the setting of the temperature. Use <> <> to set the temperature. Use <F4>/[ ⇅ ] to select the setting of the concentration. 16. Set the concentration of the standard solution with <> <> and press <ENTER>. The standard solution is measured. The measured value is checked for stability (AutoRead). 01.08.2015 08:00 17. Wait for the end of the AutoRead measurement or accept the calibration value with <ENTER>. The calibration display for the next standard solution appears. Standard 01.08.2015 08:00 18. Continuing with three- to seven-point calibration Press <ENTER> to continue with three-point calibration. or Finish the calibration procedure as a two-point calibration with <M>. The new calibration values are displayed. Repeat the steps 8 to 13 in the same way with the third and further standard solutions as necessary. The new calibration values are displayed after the last calibration step was completed. Based on the calibration data, the calibration curve is determined in sections, according to the Nernst equation modified by Nikolski. 52 ba76142e04 09/2015