1.11 DolphiCam Settings & Functionality 1.11.1 Start Gate and End Gate Setting your gate enables you to study a specific depth of your material. The start and end gate can be set between -0.38 mm and 8.8 mm (CF08) or 18.38mm (CF16). The gates can be set in different ways: · Entering a specific depth in the Start Gate and End Gate fields respectively, or · Adjusting the gates using the + (plus) and – (minus) buttons on each side of the Gate Value field, or · Click the mouse in the vertical B-Scan to get your desired gate. If the cursor is closest to the Start Gate line, the Start Gate value will be adjusted. If the cursor is closest to the End Gate line, the End Gate value will be adjusted. NOTE: · The gate values will take effect in your next image. · Amplitude image and Time-of-Flight image have separate Start Gate and End Gate settings. 16