Download Method 5 Manual - Apex Instruments, Inc.
Transcript
I S O K I N E T I C S O U R C E S A M P L E R M A N U A L Source Sampler Console To inspect the inside components of the Source Sampler Console, make sure the power cord is disconnected. Remove the front panel and tray from the cabinet by turning the four panel latches (in corners of front panel) counterclockwise with a flathead screwdriver until the latch releases. Slide out the panel using the handle at the bottom-center of the faceplate. Once the panel is partially pulled out, reach inside and disconnect the fan wire from the fan assembly. Visually inspect all of the mechanical and electrical components. Clean any accumulated dust off the components. Dry Gas Meter The dry gas meter, shown in Figure 3-11, is not field adjustable. The routine maintenance consists of performing the required periodic calibrations and calibration checks. If the dry gas meter fails repeatedly to calibrate against a wet test meter, then return the meter to the factory for repair. Valve Mechanism Reverse Flow Stopper Pitot Flow Checker to monitor leakage down to 0.02 liters Reinforced Polymer Diaphragm Figure 3-11 Positive Displacement Diaphragm Type Dry Gas Meter Apex Instruments, Inc. | Phone: 919.557.7300 | Fax: 919.557.7110 | Web: www.apexinst.com | e-mail: info@apexinst.com 79