The SETTINGS menu
The settings menu gives access to the User settings and the Machine settings.
Notice that the Machine settings are password protected and can only be
changed by authorized users.
Because the User settings and Machine settings contain an extensive list of
functions, they are explaned in special chapters
3.1.6 User settings and chapter 3.1.7 Machine settings
3.1.6
User settings
The User settings menu contains the following options:
•
•
•
•
•
•
3.1.6.1
Scanning
Dynamic reference
Image
Mouse
Workpiece compensation
Probe sound
Scanning
With the K-CMM system, a scanning mode is possible. This means that
measurements are taken continuously while moving the probe tip.
Three different ways of triggering the measurements in scanning mode are
possible:
•
•
•
Continuous measurement: measures points at the maximum frequency.
Time triggered: takes trigger points with a certain time interval, which is set
by the parameter (in ms).
Distance triggered: This mode triggers measurement points at equal
distances from each other. If it is set at a distance of 50mm, trigger points
will be taken every time the probe tip is moved more than 50mm in a 3D
volume. This method always gives a number of points, independent from
the probe’s speed.
Notice: In the scanning mode you should check whether the measurement
software and the selected interface protocol can keep up with the processing of
the amount of measurement data.
Pressing the Function button on the probe activates the scanning mode.
Therefore this button must be configured for the scanning mode (see Function
button function). When you’re in scanning mode you can start taking
measurement points by pressing the Trigger button. To stop taking measurement
points, press the trigger button again.
When the scanning mode is active, the red light in the Function button lights up
and the Mode window in the K-CMM main screen will say ‘Scanning mode’.
KRYPTON Help Pages on K-CMM