Operation
Calibration with calibration solution
(CAL 1100)
Be sure to use known calibration
solutions and the respective temperature-corrected table values (see
Calibration Solutions, Pg. 33).
Note
CAL
Activate calibration by pressing the
CAL key.
Using the Y , " keys enter
mode code “1100” and then press
ENTER.
Immerse the sensor in the calibration solution.
After approx. 6 sec the lower display alternately shows the conductivity and temperature values. Read
the conductivity value corresponding to the displayed temperature
from the table of the used calibration solution (for tables see Pg. 33).
Using the Y , " keys change the
cell factor until the display shows
the conductivity value from the
table.
Make sure that the temperature is
stable during the calibration procedure.
22
ENTER
Press ENTER to confirm the cell
factor.
The Transmitter remains in the Hold
state. You can end the Hold state
with ENTER. After 20 sec (measured value stabilization) the Transmitter returns to measuring mode.