Download Model 421 - Lake Shore Cryotronics, Inc.
Transcript
Lake Shore Model 421 Gaussmeter User’s Manual ALARM AND RELAY (Continued) The Model 421 has an audible alarm annunciator or beeper. The beeper will sound when the instrument is in an active alarm state. If the sound of the beeper is not appropriate for your application, it can be turned on or off by the user. To turn the beeper on or off, press and hold the Alarm key until the following message appears. Select With Audible ¡On °® Off Use the s or t keys to cycle between audible alarm On or Off. Press Enter to accept the new value. When the alarm feature is used to detect an error condition an active alarm state is expected when the field reading is higher than the high alarm setpoint or below the low alarm setpoint. This operation is achieved by setting the inside/outside parameter to outside. It is called outside because the alarm is active when the reading is outside the range of the two setpoints. During magnet testing or sorting it is often desirable to have the alarm active when the field reading is inside or between the two setpoints. This operation is achieved by setting the inside/outside parameter to inside. (see examples below) To select inside/outside alarm operation, continue from Audible On/Off screen or press and hold the Alarm key until the Audible setting window appears, then press Enter until the following message appears. Select With Alarm ¡In °® Out Use the s or t keys to cycle between the alarm triggered inside (In) or outside (Out) alarm setpoints. Press Enter to accept the changes. Model 421 can be configured to display pass or fail when it is used during repetitive magnet testing or sorting operations. The sort message can be turned on or off as necessary and does not affect other operations of the alarm feature. When sort is turned on the lower display line will show: Fail Low Field below low alarm setpoint. ** Pass ** Field between the two alarm setpoints. Fail High Field above high alarm setpoint. The message will use the lower display line during max hold or relative measurements. The live reading will be replaced during max hold operation and the relative setpoint will be replaced during relative operation. To enable the sort message continue from Inside/Outside or press and hold the Alarm key until the Audible setting window appears then press the Enter key until the following message appears. Select With Sort ¡On °® Off Use the s or t keys to select between Sort On and Off, then press Enter to accept the new setting. Operation 3-9