Lake Shore Model 421 Gaussmeter User’s Manual
ALARM AND RELAY (Continued)
The Model 421 has an audible alarm annunciator or beeper. The beeper will sound when the
instrument is in an active alarm state. If the sound of the beeper is not appropriate for your application,
it can be turned on or off by the user. To turn the beeper on or off, press and hold the Alarm key until
the following message appears.
Select With
Audible ¡On
°®
Off
Use the s or t keys to cycle between audible alarm On or Off. Press Enter to accept the new value.
When the alarm feature is used to detect an error condition an active alarm state is expected when the
field reading is higher than the high alarm setpoint or below the low alarm setpoint. This operation is
achieved by setting the inside/outside parameter to outside. It is called outside because the alarm is
active when the reading is outside the range of the two setpoints. During magnet testing or sorting it is
often desirable to have the alarm active when the field reading is inside or between the two setpoints.
This operation is achieved by setting the inside/outside parameter to inside. (see examples below)
To select inside/outside alarm operation, continue from Audible On/Off screen or press and hold the
Alarm key until the Audible setting window appears, then press Enter until the following message
appears.
Select With
Alarm
¡In
°®
Out
Use the s or t keys to cycle between the alarm triggered inside (In) or outside (Out) alarm setpoints.
Press Enter to accept the changes.
Model 421 can be configured to display pass or fail when it is used during repetitive magnet testing or
sorting operations. The sort message can be turned on or off as necessary and does not affect other
operations of the alarm feature. When sort is turned on the lower display line will show:
Fail Low
Field below low alarm setpoint.
** Pass **
Field between the two alarm setpoints.
Fail High
Field above high alarm setpoint.
The message will use the lower display line during max hold or relative measurements. The live reading
will be replaced during max hold operation and the relative setpoint will be replaced during relative
operation.
To enable the sort message continue from Inside/Outside or press and hold the Alarm key until the
Audible setting window appears then press the Enter key until the following message appears.
Select With
Sort
¡On
°®
Off
Use the s or t keys to select between Sort On and Off, then press Enter to accept the new setting.
Operation