13.4 Message Handling
A message is defined to begin with an attention character and end with a
valid terminator. There is no time limit between characters. Whenever an
attention character is detected, any current characters in a device receive
buffer are flushed and a new message handling sequence is begun.
Whenever a terminator is detected, the current receive buffer is passed on
to the message parser. If the terminator is not received and the controller’s
receive buffer becomes full, additional characters are ignored until a
terminator (or an attention character) is received.
13.5 ASCII Command Set Definition
The following template is used to define the controller’s commands, queries
and responses:
Description:
Verbose description of the purpose of the command
or query (message).
Keyword:
The actual ASCII characters comprising the
command or query field of the message.
Type:
‘Command’ - message item supports the
command format. ‘Query’ - message item supports
the query format.
Parameters:
A verbose description of the associated data
required to complete the message. This information
comprises the parameter field in the message.
Parameters may be ASCII numerals representing
numbers, or ASCII alpha-numeric characters,
representing a phrase.
Limits:
The allowable limits of the parameter values.
If the parameters are numeric, then the
minimum and maximum values are listed. If the
parameters are phrases, then the set of allowable
phrases are listed.
Reset Value:
If the message represents a setup or control value,
the factory default value for the message is
specified here.
Query Response:
If the message supports a query format, the
controller’s allowable responses are listed here.
Related keywords;
A list of related commands or queries for reference.
Examples:
Example uses of the command or query are
listed here.
A.4
959 Hot Cathode Controller