Download PXI E Series User Manual - Data Acquisition Unit. TJ
Transcript
Appendix A Specifications for PXI-6040E, PXI-6070E, and PXI-6071E Triggers Analog Trigger Source PXI-6040E and PXI-6070E.............ACH<0..15>, external trigger (PFI0/TRIG1) PXI-6071E.......................................ACH<0..63>, external trigger (PFI0/TRIG1) Level .......................................................± full-scale, internal; ±10 V, external Slope .......................................................Positive or negative (software selectable) Resolution ...............................................8 bits, 1 in 256 Hysteresis................................................Programmable Bandwidth (–3 dB) PXI-6070E and PXI-6071E.............2 MHz internal, 7 MHz external PXI-6040E.......................................650 kHz internal, 3.0 MHz external External input (PFI0/TRIG1) Impedance........................................10 kΩ Coupling ..........................................DC Protection.........................................–0.5 to Vcc + 0.5 V when configured as a digital signal, ±35 V when configured as an analog trigger signal or disabled, ±35 V powered off Digital Trigger Compatibility ..........................................TTL Response .................................................Rising or falling edge Pulse width .............................................10 ns min PXI E Series User Manual A-12 © National Instruments Corporation