Installation Site Requirements
Safety Hazards of Strong Magnetic Fields
The potential safety hazards of strong magnetic fields to devices such as certain
pacemakers must be understood and planned for. A set of plots indicating the
magnitude of the stray fields for each type of magnet is included in Appendix B on
page 41. These plots show typical levels of stray field. Actual levels may vary and
should be checked after a particular magnet has been installed.
Cardiac pacemaker wearers should refrain from entering a zone that would subject a
cardiac pacemaker to a magnetic intensity that could cause adverse effects. In some
instances, this zone might include space on the floors directly above and below the
magnet. Figure 4 indicates the typical 5-gauss level zone for each magnet bore size. For
assistance in determining the effect of a system on pacemaker, consult the pacemaker
user's manual, contact the manufacturer, or confer with a physician to determine the
effect on a specific pacemaker. Actual levels vary and should be checked after a
particular magnet has been installed.
Varian provides signs warning of magnetic field hazards. Refer to Appendix C on page
50 for an explanation of the types of signs and the sign posting requirements.
Additional signs are available from Varian at no charge.
750/51 6.04 m
600/51 3.60 m
0m
400
300 /5
/54
.05 m 5
00/51
9 3
0/8
2.7
40
5m
0m
200/8
2
.
2
9
2
4
0 m 200
/54 .
1.7
Magnet / Bore
(MHz/mm)
/89 2.20 m
300
7m
1.4
Magnet
center
5-gauss field
(m)
Figure 4. Typical radial distances at which 5-gauss levels exist
10
UNITYplus Installation Planning Guide
87-195320-00 G0994