Download 7312 Users Guide.vp

Transcript
15 Troubleshooting
Problem
Causes and Solutions
The controller controls or attempts to
control at an inaccurate temperature
If the controller appears to operate normally except that the bath’s
temperature does not agree with the temperature measured by the
user’s reference thermometer to within the specified accuracy, consider the following:
Erroneous parameters. Check that the calibration parameters are
all correct according to the Report of Test. If not, reprogram the
constants. If the controller does not keep the correct parameters,
the memory backup battery may be weak causing errors in data.
See “Low Battery” in a previous solution.
Poor uniformity. There may be an actual difference between the
bath’s control probe and the reference thermometer due to excess
gradients in the bath. Check that the bath has an adequate amount
of fluid in the tank and that the stirrer is operating properly. Also
check that the reference thermometer and control probe are both
fully inserted into the bath to minimize temperature gradient errors.
Defective control probe. Check that the control probe has not
been struck, bent, or damaged. Refer to the previous solution for
how to check the probe’s resistance.
The controller shows
that it is controlling at
the proper temperature, but the bath
temperature is unstable
If the bath does not achieve the expected degree of temperature
stability when measured using a thermometer, consider the following:
Wrong proportional band setting. If the proportional band is set
too narrow, the bath will oscillate causing poor stability. In this
case, increase the width of the proportional band.
If the proportional band setting is too wide, the long-term stability of
the bath is affected. In this case decrease the width of the band.
(Refer to Section 10.7.)
Bath fluid is too thick. Make sure that the bath fluid used is less
than 50 centiStokes (10 is ideal) at the temperature at which the
bath is controlling. Check the fluid manufacturer’s specifications.
You should also change the bath fluid regularly and if it changes
colors or becomes too thick.
Defective control probe. Check that the control probe has not
been struck, bent, or damaged. Refer to the previous solution for
how to check the probe’s resistance.
The controller alternately heats for a
while then cools
Wrong proportional band setting. If the proportional band is set
too narrow, the bath will oscillate between too much heating and
too much cooling causing instability. Increase the width of the proportional band until the temperature stabilizes. (Refer to Section
10.7.)
The bath does not
Insufficient cooling. This may be caused by lack of refrigerant beachieve low tempera- cause of a leak in the system. Refer to Section 13, Charging Intures
structions.
74
Manual Rev. 152302
7312