Transcript
15 Troubleshooting Problem Causes and Solutions The controller controls or attempts to control at an inaccurate temperature If the controller appears to operate normally except that the bath’s temperature does not agree with the temperature measured by the user’s reference thermometer to within the specified accuracy, consider the following: Erroneous parameters. Check that the calibration parameters are all correct according to the Report of Test. If not, reprogram the constants. If the controller does not keep the correct parameters, the memory backup battery may be weak causing errors in data. See “Low Battery” in a previous solution. Poor uniformity. There may be an actual difference between the bath’s control probe and the reference thermometer due to excess gradients in the bath. Check that the bath has an adequate amount of fluid in the tank and that the stirrer is operating properly. Also check that the reference thermometer and control probe are both fully inserted into the bath to minimize temperature gradient errors. Defective control probe. Check that the control probe has not been struck, bent, or damaged. Refer to the previous solution for how to check the probe’s resistance. The controller shows that it is controlling at the proper temperature, but the bath temperature is unstable If the bath does not achieve the expected degree of temperature stability when measured using a thermometer, consider the following: Wrong proportional band setting. If the proportional band is set too narrow, the bath will oscillate causing poor stability. In this case, increase the width of the proportional band. If the proportional band setting is too wide, the long-term stability of the bath is affected. In this case decrease the width of the band. (Refer to Section 10.7.) Bath fluid is too thick. Make sure that the bath fluid used is less than 50 centiStokes (10 is ideal) at the temperature at which the bath is controlling. Check the fluid manufacturer’s specifications. You should also change the bath fluid regularly and if it changes colors or becomes too thick. Defective control probe. Check that the control probe has not been struck, bent, or damaged. Refer to the previous solution for how to check the probe’s resistance. The controller alternately heats for a while then cools Wrong proportional band setting. If the proportional band is set too narrow, the bath will oscillate between too much heating and too much cooling causing instability. Increase the width of the proportional band until the temperature stabilizes. (Refer to Section 10.7.) The bath does not Insufficient cooling. This may be caused by lack of refrigerant beachieve low tempera- cause of a leak in the system. Refer to Section 13, Charging Intures structions. 74 Manual Rev. 152302 7312