Achieva Ventilator Clinician’s Manual
Operation
Assist/Control mode, volume breaths
To operate the ventilator in Assist/Control mode, delivering breaths of a
selected volume:
1. Press MENU/ESC to enter the ventilator’s menu system.
2. Set alarm latching (see Latching and non-latching on page 6-4),
or pressure trigger level (see Pressure trigger on page 5-5) and/or
operating altitude (see Altitude setting on page 5-6) from the
menu, as needed.
3. From standby, press Start/Enter (
) to display the parame-
ters.
4. Set MODE to A/C.
5. Set the prescribed volume.
6. Set the prescribed inspiratory time.
7. Set the inspiratory sensitivity to an appropriate level.
8. Set the breath rate.
9. Set Pressure to 0.
10. Set PEEP as required.
11. Set the low pressure and high pressure alarm levels. See Setting the
low pressure alarm on page 4-27.
12. If required, set FIO2 (available only on Achieva PSO2).
13. Press Ventilate (
November 2003
) to begin breath delivery.
5-7