Download Sedex 85LT Light Scattering Detector manual
Transcript
Installation of the Detector RIGHT WRONG Figure 2-6: Fixing the Glass Chamber c) Use the large black nut to position the nebulizer on the glass chamber. d) Screw the inlet fitting in the bulkhead on the left side of the compartment. Special care must be taken when positioning this fitting. The nebulizer is terminated with a small piece of Teflon tubing with an outer green sleeve. For proper operation, the Teflon tubing must extend less than 2mm past the end of the green sleeve (Figure 2-7). Figure 2-7: Nebulizer Inlet Fitting Fill the siphon overflow on the nebulizer/glass chamber assembly with the mobile phase that will be used for the separation. If you are using a very volatile solvent (e.g. hexane or CH2Cl2), then use water to fill the overflow. The liquid should fill the bent part of the siphon, but should not pool in the bottom of the evaporation tube. Make sure that no liquid leak could affect the detector performance or create laboratory pollution. 2.5.6 Connecting the Siphon Overflow Attach the drain tube assembly to the end of the siphon tube using the tapered hose connector and lead the tube to waste and drain. Locate the tube in such a way that the discarded part of the solvent can flow freely from the siphon and ensure that the end of the tube never dives in the collected liquid. Make sure that the liquid waste container complies with the solvent nature. Note: A drain tube with a bend or diving in the liquid will create pressure fluctuations in the detector and will result in bad measurement results. SEDEX Model 85LT Evaporative Light-Scattering Detector 2-13