DX......distance tracking with target board (50Hz)
Input parameter
Output
SA, SD, SE, SF, OF
RS232/RS422, digital switching output, analog output
DX mode performs at a steady measuring rate of 50 Hz (only LDM42).
As a necessary precondition for measured values to be stable, a white
target board must be placed at the selected object.
This measurement mode is intended in the first place for objects
performing homogeneous motion up to 4m/s. For higher rates of
measurement, preceding measured values will be included in the
process to calculate a currently measured value. There must be no
sudden jumps in distance greater than 16 cm within the measuring field!
Caution: To prevent transmission problems, please use
9600 baud as the minimal baud rate for DX Mode.
8.6.
DF......distance measurement with external trigger
Input parameter
Output
18
SD, SE, SF, ST, OF, TD, ST, SA, RM
RS232/RS422, digital switching output, analog output
DF mode allows a measurement that is triggered by an external trigger
pulse.
Initially, after selecting this mode, the operator does not receive any
response. As soon as the trigger pulse has been detected, the
LDM 41/42 A will send data and switches to digital and/or analog output.
Settings for trigger delay (delay) and trigger slope can be defined via
parameter TD (see 8.22 TDx y......display/set trigger delay and trigger
level).
With ST=1 the maximum trigger frequency should not exceed 3.5 Hz.
18
Trigger function is not available for devices with internal heating (LDM 41/42 A /h)