SonTek/YSI
3.5.3. Data Recording
The real-time software records Argonaut data in a standard binary file format.
• The same format is used whether data are recorded using the real-time data collection
software or the internal recorder.
• Within the standard data file format, there are two options for the format of individual
samples: LONG or SHORT. The SHORT format is for autonomous deployments to reduce
storage requirements for long deployments. Real-time data collection always uses the
LONG data format.
• When data collection is started, the hardware configuration and user operating parameters
are stored in a file header structure.
• With each sample (using the LONG format), the system stores a record containing time,
velocity, signal strength, standard deviation, heading, pitch, roll, temperature, pressure,
and several other parameters.
• All the data contained in these files can be accessed by the data conversion programs
discussed in Section Section 6.
• The binary file format is described in detail in the Argonaut Operation Manual.
When using the real-time software to collect data, you also have the option of storing data
internally on the Argonaut recorder.
• The real-time software does not affect settings that relate to the internal recorder; these are
left as they were when the software was started.
• The most common reason to use the internal recorder while collecting real-time data is for
a backup in case of computer malfunction or power loss.
• To use the internal recorder as a backup, do the following.
○ Run our SonTerm terminal emulator program (see §4.2-Windows; Section 9-DOS).
○ Send a BREAK to establish communication.
○ Erase the internal recorder if necessary using the Format command. Be sure to first
retrieve and back up any files currently stored on the recorder.
○ Send the command Recorder ON.
○ Set a deployment name using the Deployment <name> command.
○ Send the command SaveSetup.
○ You can now run the real-time software, and the Argonaut will also record data on the
internal recorder.
Argonaut Software Manual (March 1, 2001)
25