carefully and evenly (this is not necessary when using disposable electrodes with built-in gel). The following figure shows the standard positioning of the electrodes. Standard electrodes positioning of the V1: on the 4th intercostal space, right parasternal; V2: on the 4th intercostal space, left parasternal; V3: on the 5th rib, between V2 and V4; V4: on the 5th intercostal space, left hemiclavicular; V5: on the left anterior axillary, same level as V4; V6: on the left mid-axillary at the level of V4; Peripheral electrodes: generally a few centimetres above ankles and wrists. Warning: make certain that the conductive parts of the electrodes are not in contact one with another or with other metallic parts. In any event, silver and silver chloride electrodes are designed and manufactured in such away as to minimise the likelihood of accidental contact between conductive parts and external metal objects. Ensure that the device is not affected by disturbances originating from the AC mains power supply (see “Initial preparation”). 4.3 Select recording characteristics operating mode, display and print format, speed, sensitivity, filter Operating mode The available recording modes (Operating mode) depend on the active configuration of the electrocardiograph. to select the desired mode; the corresponding choice is displayed on the screen. The options available are: Automatic, Manual, Personalized. To change Personalize Mode see "Settings". During printing, the active mode is printed on the information line of the printout. 30